<link rel="stylesheet" href="styles.f3b1fba60ec7970c.css">
Nasarawa State University, KeffiNasarawa State University, KeffiInstitutional Repository

DETERMINATION OF ABSORBED DOSE IN COMPUTED TOMOGRAPHY EXAMINATIONS OF ADULT PATIENTS AT FEDERAL MEDICAL CENTRE, KEFFI, NIGERIA

dc.contributor.authorJoseph, Gabriel Isibor
dc.date.accessioned2021-04-22T11:41:01Z
dc.date.available2021-04-22T11:41:01Z
dc.date.issued2017-07-21
dc.descriptionDissertationen_US
dc.description.abstractThe use of Computed Tomography (CT) in medical diagnosis delivers radiation doses to patients that are higher than those from other radiological procedures. The aim of this study was to measure radiation dose in CT examination of adult patients in Federal Medical Center, keffi. Details were obtained from approximately 38 CT examinations carried out at Federal Medical Center, Keffi. The CT examination includes head, chest and abdomen. The weighted CT dose index (CTDIw), dose length product (DLP), and effective dose (E) were calculated from the patients detail collected from the center which includes; scan length (L), pitch (P), volume CT dose index (CTDIv). The mean value of CTDIv of the patients that underwent CT examination for Chest, Head, and Abdomen are 14.62mGy, 58.28mGy, and 15.06mGy respectively. These are found to be below the Australian reference limit which is 15mGy, 60mGy, and 15mGy for chest, head, and abdomen respectively except for abdomen which is slightly above the limit with 0.6mGy. The mean value of DLP where found to be 245.67mGy.cm, 428.12mGy.cm, and 861.89mGy.cm for head, chest and abdomen respectively. This is found to be below the reference limit of 1000mGy.cm, while that of chest and abdomen are found to be above the reference limit which are 400mGy.cm, and 700mGy.cm for chest, and abdomen respectively. There is need for the center to review and optimize the DLP setting in the machine. However, in order to achieve the above optimization strategies, there is a great demand to educate CT personnel on the effects of scan parameter settings on radiation dose to patients and image quality required for accurate diagnosis.en_US
dc.identifier.citationA DISSERTATION SUBMITTED TO THE SCHOOL OF POSTGRADUATE STUDIES, NASARAWA STATE UNIVERSITY KEFFI, IN PARTIAL FULFILMENT OF THE REQUIREMENTS FOR THE AWARD OF DEGREE OF MASTERS IN RADIATION AND MEDICAL PHYSICSen_US
dc.identifier.urihttp://localhost:8080/xmlui/handle/123456789/1173
dc.language.isoenen_US
dc.publisherDEPARTMENT OF PHYSICS FACULTY OF NATURAL AND APPLIED SCIENCES NASARAWA STATE UNIVERSITY, KEFFI NIGERIAen_US
dc.titleDETERMINATION OF ABSORBED DOSE IN COMPUTED TOMOGRAPHY EXAMINATIONS OF ADULT PATIENTS AT FEDERAL MEDICAL CENTRE, KEFFI, NIGERIAen_US
dc.typeThesisen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
DETERMINATION OF ABSORBED DOSE IN COMPUTED TOMOGRAPHY EXAMINATIONS OF ADULT PATIENTS AT FEDERAL MEDICAL CENTRE, KEFFI, NIGERIA.pdf
Size:
834.65 KB
Format:
Adobe Portable Document Format
Description:
Dissertation

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: