Structural Shielding Evaluation: A Case Study of the Radiography Room of a Rural Hospital in Jos, Nigeria
Loading...
Date
Journal Title
Journal ISSN
Volume Title
Publisher
Department of Physics, Nasarawa State University Keffi
DOI
Abstract
Introduction: Most Nigerian rural hospitals constructed before the
publication of the 2005 National Council on Radiation Protection
and Measurements Report No. 147 did not undergo a performance
assessment at the time of construction. To avoid overexposure to
ionizing radiation and to ensure adequate protection of patients,
workers, and the public, the shielding barriers need to be evaluated
to ascertain that they conform to this standard.
Methods: This study evaluates the shielding barriers for the general
radiography room in a rural hospital in Jos, Nigeria. The workload
information, generator voltage waveform, anode material, filtration,
and anode angle with XRAYBARR calculation model were used to
estimate the thickness of lead, concrete, gypsum, steel, plate glass,
and wood required to shield the X-ray facility installed in the hospital.
The design dose limit was compared to the estimated shielded
dose, and the calculated shielded barrier thickness to the design
shielded barrier thickness was also compared.
Results: The unshielded radiation doses inside the X-ray room were
high, indicating that the radiological department of the study area is
not minimizing radiation doses to patients. The calculated doses
beyond the barriers were greater than the design dose limit, indicating
that the shielded barriers in place were not adequate and
did not comply with the international standard.
Discussion: Hospitals must understand the type of shielding materials
that can provide adequate protection and to what extent they
can protect their radiography rooms. Management and radiation
protection agencies need to ascertain whether these barriers are still adequate or require reinforcement through regular quality assurance
testing. Due to an increase in workload associated with an increase in
population and urbanization, proper policies are needed more than
ever in this and other rural hospitals in Nigeria. Conclusion: The tested X-ray rooms did not comply with international
recommendations for shielding thickness. Except for the
door, console, and changing room, a 0.5 mm-thick lead reinforcement
is required. It is also recommended that quality assurance
testing occur on an annual basis.
Description
Citation
[1] Maliro, J.F.(2011).Ionizingradiationprotectionawarenessamong nurses workingatQueenElizabethcentralhospitalinMalawi.PhD Thesis (pp.20–25).SouthAfrica:UniversityofJohannesburg. [2] International CommissiononRadiologicalProtection(1991).1990 RecommendationsoftheICRP(pp.1–3).NewYork:ICRPPublication 60 AnnalsoftheICRP21.