ASSESSMENT OF DIAGNOSTIC REFERENCE LEVELS FOR DOSE OPTIMIZATION IN COMPUTED TOMOGRAPHY (CT) IN SELECTED HOSPITALS IN PLATEAU AND NASARAWA STATES, NIGERIA
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Department of Physics, Nasarawa State University Keffi
DOI
Abstract
This study assessed the diagnostic reference levels in computed tomography in some selected hospitals
in Plateau and Nasarawa States, A retrospective study was conducted in two high traffic diagnostic
hospitals in the two States with the sample size of 167 adult patients were considered for this project,
their dose data of computed dose tomography index (CTDIv) and dose length product (DIP) were
collected and the mean CTDIv, DLP, standard deviation, minimum, maximum and the 75th percentile of
the dose were evaluated using SPSS software. Centre A: has 64 number of slices with 91 sample size and
the scanner type is Philips. Center B: has 16 number of slices with 76 sample size and the scanner type is
GE. The routine CT examination of head, chest and abdomen were reviewed and the DRLs for each body
region defined as the 75th percentile of its CTDIv and DLP. In Centre A, the total number of 91 patients
were used for the study and comprising of 50 males and 41 females across the body regions, representing
a gender distribution of Head (55.7% and 44.3), chest (53.6% and 46.4%) and abdomen (53.5% and 46.5%)
for male and female respectively. The mean and standard deviation values of the ages are: Male 52.9 and
female 14.7 for head, male 46.9 and female 13.9 for Chest, male 57.4 and Female 13.9 for abdomen
respectively. However, the established facility reference levels (FDRLs) for centres A and B are: 46.4mGv,
990.5mGy.cm,14.2mGy/526.7mGy.cm,14.0mGy/710.7mGy.cm;48.5mGy/1027.8mGy.cm,13.4mGy/525.2
, 9.9mGy/537.8mGy.cm, for head, chest and abdomen respectively. Findings from this study for DRLs in
terms of CTDIv of 46.4 mGy, 14.2mGy and 14 for centre A, 48.5 mGy, 13.4 mGy and 9.9mGy for centre
B show that they are in line with the EC. (1999) and similar to the findings by (Sadri et al.. 2010). (Samos
et cil., 2013) and (Karita el al., 2009). (Wanbani et al., 2010) findings of 1364, 745 and 1143mGy.cm are
as high as the findings of this study of 1027.8, 525.2 and 537.8mGy.cin and was higher than the EC. <
recommended values of 1050, 650 and 900mGy.cm. The DRLs in terms of DLP findings from this stud}
are in line with findings used in this literature, but there are little difference of some values which call to
action to follow the existing scan protocols and optimization of practice are strongly advised. When
procuring CT scan machines, qualified medical physicists should be consulted to advice on the trend and
dose saving scanners. Like Siemens CARE Dose 4D AEC, that has an inbuilt dose saving mechanism. The
Hospitals should engage a medical physics service that offers clinical dose optimization service to help
hospitals manage patient doses and to review imaging protocols.